Bispectral Index Monitoring Documents Burst Suppression During Pentobarbital Coma
Vice-Chairman, Dept of Anesthesiology, Chief, Division of Pediatric Anesthesiology
* To whom correspondence should be addressed. E-mail: tobiasj{at}health.missouri.edu.
2, 3-5, and 6 bursts/min, respectively. The sensitivity and specificity of a bispectral index value of 10 to 20 were 96% and 92%, respectively, whereas the sensitivity and specificity of a suppression ratio of 65% to 85% were 89% and 88%, respectively, in demonstrating the presence of 3 to 5 bursts/min. Bispectral index monitoring may be easier to perform and may require less technical expertise to interpret.
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
2, 3-5, and
6 bursts/min, respectively. The sensitivity and specificity of a bispectral index value of 10 to 20 were 96% and 92%, respectively, whereas the sensitivity and specificity of a suppression ratio of 65% to 85% were 89% and 88%, respectively, in demonstrating the presence of 3 to 5 bursts/min. Bispectral index monitoring may be easier to perform and may require less technical expertise to interpret.